For different types of liquids, pipettes have different cleaning and maintenance methods

For different types of liquids, pipettes have different cleaning and maintenance methods

Liquid to be processed Prevention method Cleaning and maintenance methods
 Aqueous solution and buffer  Disassemble the lower end of the pipette, rinse the contaminated parts with double distilled water, dry under 60℃ or air dry, and apply a small amount of grease to the piston
 Inorganic acid or base  =Use pipette tip with filter  After pipetting, remove the lower end of the pipette and place it in a ventilated place. If necessary, use double distilled water to rinse the contaminated parts, dry them below 60℃ or air dry, and apply a small amount of grease to the piston
 Infectious fluid  =Use pipette tip with  filter = Forward pipetting  = Forward pipetting Remove the lower end of the pipette, soak it in the laboratory’s conventional disinfectant for 30 minutes, then wash it with double distilled water, and autoclave the contaminated parts at 121°C, 1 bar, and 20 minutes
 Cell culture  =Use pipette tip with filter   = Forward pipetting Remove the lower end of the pipette, soak it in the laboratory’s conventional disinfectant for 30 minutes, then wash it with double distilled water, and autoclave the contaminated parts at 121°C, 1 bar, and 20 minutes
 Organic solvents  = Forward pipetting   After pipetting, remove the lower end of the pipette and place it in a ventilated place. If necessary, use double distilled water to rinse the contaminated parts, dry them below 60℃ or air dry, and apply a small amount of grease to the piston
 Radioactive solution   =Use pipette tip with filter = Forward pipetting   Remove the lower end of the pipette, soak the contaminated parts in a composite solution or a special cleaning solution for 30 minutes, then wash with double distilled water, dry at 60℃ or air dry, and apply a small amount of lubrication to the piston fat.
 Nucleic acid or protein solution   =Use pipette tip with filter = Forward pipetting  Nucleic acid: Boil in aminoacetic acid or hydrochloric acid buffer (pH2.0) for 10 minutes, rinse with double distilled water, dry at 60℃ or dry naturally, and apply a small amount of grease to the pistonProtein: Remove the lower end of the pipette, clean it with detergent, rinse with double distilled water, dry at 60℃ or dry naturally

For different types of liquids, pipettes have different cleaning and maintenance methods


Post time: Oct-27-2021